| Weight | 69156723416 kg |
|---|---|
| Dimensions | 6923787706 × 6923557705 cm |
No features added.
| Input Power | Operating Voltage: 180-265 Volt AC, 50-60 Hz Max. Power input: 1000 VA, Fuse: 6 A, UPS Backup : 1KVA |
|---|---|
| Dimension | L x W x H in cm - 38 x 31 x 12, Weight in kg < 6 |
| MONOPOLAR CUT | Pure Sine wave 390 kHz Low - Continues, 300 W at 300 £2, CF 1.5, Low Volt Pure - Continues, 300 W at 300 Q, CF 1.5, High Volt Blend - Bursts 29 kHz, 50%, 200 W at 300 Q, CF 2.5 |
| BIPOLAR CUT | Pure Sine wave 390 kHz BiCut - Higher voltage, 70 W at 100 Q, CF 1.5 |
| LEAKAGE CURRENTS | • RF leakage current ⚫Monopolar: < 150 mA rms • Bipolar: < 60 mA rms • Low Frequency leakage currents • Normal polarity, intact ground: < 10 μA • Normal polarity, ground open: < 50 μA |
| FEEDBACK SYSTEM | • 3 Microprocessor Controlled 6SENSE Technology with sampling rate above 4,00,000 sense/second • Maximum output power deviation < 15% or 5 watts, whichever is greater at any setting with PER-98 |
| MONOPOLAR COAG | Damped Sine wave 460 kHz Soft - Continues at 83 kHz, 120 W at 500 Q, CF 3.9, Fulgurate - Repetition Frequency 29 kHz, 120 W at 500 Q, CF 7.0 Spray - Randomized Repetition 25 kHz<f<50kHz, 120W at 500 Q2, CF 8.3 |
| BIPOLAR COAG | Pure Sine wave 390 kHz Micro - Lower voltage, 70 W at 100 Q, CF 1.5 Standard - Medium voltage, 70 W at 100 Q, CF 1.5 Force - Higher voltage, 70 W at 100 Q, CF 1.5 |
| PREM SYSTEM | Measuring Frequency: 80 kHz ± 10 kHz. Measuring current : < 10 μA Acceptable Resistance Range Dual Area PREM Patient Return Electrode : 5-135 Q (Adaptive) Single Area Patient Return Electrode: < 10Ω |
| DESIGN STANDARD | Classification according to the EC- Directive 93/42/EEC: Class Ilb General Safety Standards: IEC 60601-1-1, IEC 60601-2-2 EMI/EMC Compatibility Standard: IEC 60601-1-2 IS Safety Standard: IS 7583 |